OEM 스펙트로메타 - SDC

(SPECTRA TO DIGITRAL CONVERTER (SDC-VIS))

 

Spectra to Digital Converter (SDC-VIS)

OEM Spectrometer

ATX´s new spectraprocessor SDC is the first device combining an optical bench (flat field imaging spectrograph) with the read out/ processing electronics in a small dual in line 24 pin housing. Therefore it can be used like a standard electronic IC with an optical input. The data access takes place similarly to a memory device (e.g. RAM).

 

Description

  • internal storage of wavelength - pixel relation
  • storage of reference data
  • preprocessing of data, dark signal correction, absolute calibration
  • internal calculations like averaging, interpolation, transmission, reflexion, absorption, color-coordinates

 

Interfaces

  • serial interface (SPI)
  • parallel 8bit interface
  • simple UART
  • busy output
  • power down input
  • external trigger-input

 

Applications

  • mobile spectrometric measurement equipment
  • colorimeter
  • spectroradiometer
  • analytical instrumentation
  • multichannel spectrometric systems


spectrometer

 

Advantages

·

easy to implement in customer specific designs

·

processing of spectral calculations

·

customer specific algorithms can be implemented

·

simple multichannel arrangement


Specifications

Applications

 Single and multichannel spectral measuremen

Optical parameters

Spectral range

 

380 nm ... 760 nm (optional 250 nm ... 500 nm)
Other spectral ranges on request

Optical resolution>

9 nm FWHM (others on request)

Optical Input

Optical fiber, 100 µm core diameter, ST (optional SMA) terminated

Spectral resolution

9 nm (6 nm with 50 µm fiber) FWHM

Digital spectral resolution

4.1 nm

Wavelength accuracy

± 1nm (ASTM E275, filter BG 20, 2 mm, at 528.7 nm/ 684.3 nm)

Wavelength reproducibility

± 0.1 nm (ASTM E275, filter BG 20, 2 mm, at 528.7nm/ 684.3 nm)

Stray light

 

< 10-3 (ASTM E387, GG 495, 2 mm, at 430 nm)

Sensitivity

typ. 1.3 × 1014 counts/Ws (550 nm)

Spectrometer
Grating
Detector


Holographic, flat field
Photodiode array

Electronical parameters

Digital resolution

16 or 14 bit

Electronic noise

2 LSB

Integration time

4 ... 65000 ms

Power supply

5V ±5%, 100mA (5mA in power down mode)

Interfaces

serial interface (SPI)
parallel 8bit interface
UART (TTL-level)

Mechanical parameters

Dimensions

 

40,5 mm x 42 mm x 20 mm

 

Weight

 

50 g

 


   

 