 기술사양 (TECHNICAL SPECIFICATION) : OZONE ANALAZER 3-02.P-R

The range of ozone measuring concentrations

: 0 - 500 mcg/m3

Zero drift, less than

: 1 ppb / 30 days

Calibration drift, less than

: 5% / 30 days

Linearity error

: 2 % of full scale

Output signal: analogue (additionally) digital

: 0 - 5 mА or
: 4 - 20 mА
:
RS-232

Dimensions

: 145 х 270 х 390 mm

Weight, not more

: 5,5 kg

The limit of final value stabilization period То,9д , not more

: 30 sec

Power supply

: 220V, 50 Hz

Power consumption

: 15 VA

 

 

 

 운용조건 (THE AMBIENT OPERATION CONDITIONS)

Temperature

: +10 ...+35оС

Pressure

: 630 - 800 mm Hg.

Humidity

: Up to 95%

Parameters of the analyzed ambient air in the gas analyzer input :

: -

Temperature

: +10...+40оС

Volume flow

: 1,5 ± 0,2 l/min

relative humidity

: Up to 98%

 

 

 기능 (ASSIGNMENT)

Because of low dimensions and high reliability this instrument can be used for ozone background monitoring in inhabited zone, for ozone monitoring in different process working zone when concentrations are near extreme marginal values. This analyzer is a basis model of chemiluminescent ozone meters, manufactured in the enterprise. Analyzer is a main resort of measuring in the level of extreme marginal values in Russia in present.

 

 

 적용분야 (APPLICATIONS)

Industry ozone monitoring in working zone air.

 

 

 운용원리 (PRINCIPLE OF OPERATION)

Gas analyzer is a gas comparator providing high linearity of signal transformation function. The comparison is carried out between a model mixed gas from calibrator and analyzable gas. The basis of analyzer 3-02 П1 operation is an effect of the heterogeneous chemiluminescence, which is carried out due to ozone exothermic reaction with composition oxidizable chemical substances. The intensity of a composition luminescence is proportional to density of ozone in a mixed gas. Intensity is measured and transformed to the digital signal, indicated in the analyzer display. Built-in micropump provides the feed of analyzable gas sample in chemiluminescent reactor.

 

 플로우 차트 (GAS ANALYZER FLOWCHART)

 

 

 

 

포터블가스분석기, 황화수소분석기, 잔류가스분석기,오존분석기,O3, 염소분석기, 암모니아분석기, 벤젠분석기, 나프탈렌분석기,배출가스분석기, 대기가스분석기 방향족 분석기, 염기산화물분석기, 금속이온분석기, 바나듐분석기, 구리이온분석기, 수소분석기, 황산화물분석기, 색상분석기, 톨루엔분석기, 코발트분석기,CIP분석기, MEG분석기, 페놀분석기, 광화이버 스펙트로메타, 샘플링시스템, 벤젠, 자일렌, 발전소, 소각로, NOX, SOX, Process analyzers, Process stream,OZONE ANALYZER, Gas, liquid,  Hydrogen sulfide analyzer, Tail gas analyzer for the Claus sulfur recovery process, UV / VIS / NIR Diode array spectrophotometer, Chlorine analyzer, Ammonia analyzer, Aromatics analyzer, chlorine dioxide analyzer, Metal ions analyzer, MEG analyzer,  TiCl4 ,vanadium analyzer, CIP analyzer, Copper ions analyzer, Hydrogen analyzer, Sulfur dioxide analyzer, TBC analyzer, Color analyzer, Fiber optic spectrometers, Sampling systems, Cl2, BTX, Benzene, Toluene, Xylene, MEHQ, NH3, Multi Components, Concentration monitoring, On-line, Continuous measurements, Cobalt recovery, Sulfur recovery, Air demand, Ammonia slip, power plant, SO2, COS, CS2, Sulfur, H2S, NCl3, ClO2, O3"

 

Top of Page

 

 

   

주소 : 경기도 시흥시 정왕동 1288-2 번지 동우디지털파크 A-512

Tel : +82-(0)31-497-1134,   Fax : +82-(0)505-497-7710   

Url : http://www.DTX.co.kr   E-mail : aitnix@aitnix.com

All Rights Reserved, Copyright(C) DTX, INC. 2008